RF Design Magazine


In context--Internal zero and calibration for RF power sensors
Jun 1, 2005 12:00 PM  By Alan B. Anderson

RF power sensors and meters are used to add traceability and ensure accurate RF power measurements. The zero and calibration procedure must often be performed on the sensors and meters to ensure accurate measurements and involves multiple connections to an external calibration source. This article introduces Agilent's internal zero and calibration approach, which helps reduce measurement uncertainty associated with calibrating with an external source.

For the PDF version of this article, click here.

The primary use of an RF power sensor and RF power meter combination is to add traceability and accuracy to test systems and for making accurate, absolute RF power measurements. The environment may change, whether a lab test bench or a production line, either on RADAR systems, mobile phones or base stations, but in most cases, a power meter and sensor will be used as the principal route to a traceable absolute power measurement.

Central to an accurate RF power measurement is the zero and calibration procedure that the user often must perform on a sensor and meter in order to make the subsequent measurements accurate. This article will discuss the role that this process fulfills and consider how it applies to current power-sensing techniques. In particular, Agilent's ‘internal zero and calibration’ approach will be introduced, which eliminates multiple connections with an external calibration source, with the benefits of reduced measurement uncertainty, reduced connect wear and faster test times.

Sensor technology and zero and calibration requirements

Commercially available sensors rely on three types of RF detectors — thermistors, thermocouples and diodes. Each sensor type typically consists of an RF connector interface, possibly including an internal attenuator or direct current (dc) block capacitor, and the RF detector, followed by signal conditioning/processing circuitry. Figure 1 shows this for a diode-detector-based sensor.

For a thermistor detector, the detection relies on the dissipation of the incident RF power heating the detector and thus affecting a change in its resistance. By using a bridge circuit and dc substitution, the detector is making a closed loop measurement of RF power. So, in measuring the amount of substitute electrical power that must be applied to balance the RF heating effect, a measure of the power that would be dissipated in a 50 Ω load is obtained.

To improve the accuracy of this measurement, a calibration factor is applied. This accounts for mismatch loss, because the actual thermistor does not provide a perfect 50 Ω load, and also for the effective efficiency, because all the RF power is not necessarily dissipated in the thermistor itself, and therefore measured.[1] This calibration factor will vary over the RF frequency range of the sensor and is generated when the sensor is calibrated, either when it is produced, or as part of its periodic calibration cycle. It is typically constant over power levels and temperature.

So when making a power measurement with a thermistor power sensor, due to the use of the dc substitution approach, no further user calibration is required. However, a zeroing operation is required, to null any offset effects of the measurement instrumentation.

For a thermocouple and diode detector, there is no inherent capability to measure RF power prior to the devices being calibrated. These detectors are often termed ‘open-loop’ in that they rely on a calibration prior to use. They each output a voltage that is proportional to the input RF power, however, to make it meaningful, it also requires a scaling factor. This scaling factor is usually derived when the user connects the sensor to a known RF source — typically a 50 MHz, 0 dBm calibrator (also known as a reference oscillator) — thus providing a closed-loop system, and allowing the scaling factor to be calculated. This process will subsequently be referred to as a ‘user calibration.’

Consider an output of 2 mV from a thermocouple or diode power sensor. This is the result of an unknown incident power at 50 MHz being applied. The output value only becomes meaningful when the sensor is calibrated with 0 dBm, 50 MHz, which yields, say, a 1 mV output. So, by calculating the scaling factor as 1 watt per volt, and applying it to the output of 2 mV, a measurement of 2 mW is the result. Calibration factors, as introduced for the thermistor power sensor, are also applied to transfer this scaling factor to other frequencies; however, they rely on the closed loop calibration being undertaken prior to them being applied.

In addition, both diode and thermocouple sensing types also require to be zeroed prior to making a power measurement. The zero function removes residual offsets in the instrumentation when no power is applied. Typically, the user calibration and zeroing operation is performed at the same time and is often referred to as a ‘zero and cal.’

Before the calibration process can be valid, zeroing must take place. Here, when no RF power is incident on the detector, the power meter measures the output of the sensor. Continuing the example — if the zero measurement yielded 0.2 mV, the post-zeroing calibration value (previously 1 mV) should be 1 mV-0.2 mV = 0.8 mV. This value would then be used to generate a scaling factor of 0.8 watt per volt.

In the majority of power sensors the user calibration and zeroing process is applied immediately before a power measurement is made. This is because of a unique characteristic among RF test instrumentation: the power meter and sensor are separate and that the measurement plane is located away from the meter — wherever the sensor is located. The main reasons to split the system into a meter and sensor are:

  • To locate the measurement reference plane as close to the device under test (DUT) as possible. To minimize the degradation in accuracy from cabling loss and additional mismatch uncertainty from a degraded impedance match.

  • To facilitate the interchangeability of the RF sensor, providing flexibility in power range and frequency coverage at a lower cost by allowing part of the system to be interchanged.

Therefore, the sensor, cable and meter are often being used for the first time together, or have been disconnected and reconnected while assembling a test station. So for maximum accuracy, a user calibration and zeroing is carried out on the system prior to a measurement.

The evolution of zero and calibration

In modern RF power sensors[2], the detector is non-linear over part of its usable range (i.e., the detected output voltage is not always linearly proportional to input power). Additional levels of correction beyond frequency calibration factors must, therefore, be made because the calibration factors are no longer constant over power and temperature ranges.

For example, extended-dynamic range diode detectors make linearity correction over power level, which typically comprises a 50 MHz linearity characterization, which is then applied to all other frequencies. The most convenient way to combine the calibration factor correction and the linearity correction is to make all corrections relative to the frequency and power level of the RF source used for the user calibration.

Further linearity correction may also be applied at different frequencies — this correction has been termed frequency-dependent linearity correction or FDLC[3]. Also, temperature changes in linearity have been accounted for by using different linearity corrections at different temperatures. This overlaying of linearity, calibration factor and FDLC essentially means that for each temperature, frequency and power level, a unique correction factor must be applied to relate the sensor's output voltage to the input RF power. For implementation purposes these corrections are made relative to the user calibration reference source and rely on characterization data generated when the sensor is produced, or as part of its periodic calibration cycle.

The user calibration is akin to taking a global positioning system (GPS) reference for your position and using it to fix your position on a map, the knowledge of your position at one point on the map, does not ensure you know your whereabouts when you move to other locations. The only way to do this is to trust the detail of the map to provide the answer to what height you are and where you are as you move about. The GPS acts as the user calibration, and the detail on the map represents the sensor characterization — your knowledge of your position (the measurement) is only as good as your initial reference fix and the validity and how up-to-date and accurate the map is.

Traditional sensors, operating within their square-law region, represent a flat landscape of restricted area, but modern peak and average sensors with extended dynamic range represent large hills and valleys over a greater area. Hence, more recent sensors place greater importance on the accuracy of the map (the correction factors and characterization), than the initial GPS reading (the user calibration). Figure 2 shows the response behavior or ‘map’ for Agilent's P-series power sensor. The periodic calibration of the sensor is essentially a check or update of the detail of the map.

During this evolution of power sensors, a subtle change has occurred. Where once a user calibration was performed to provide the power accuracy and traceability of a power measurement. This has increasingly shifted to a reliance on the factory characterization and periodic calibration cycle to ensure accuracy.

A common misconception of the user calibration is that it can detect changes in the response of a sensor and make adjustments. However, because the sensor and meter combination may also have changed, any conclusions about a change in the RF detectors behavior cannot be made. The closed-loop also contains the connector interface, amplifier circuitry, cabling and the power meter. So the primary function of the user calibration is to combine the sensor, meter and cable, not to detect a change in the detectors behavior.

In recognition of this, the Agilent P-series N1921/2A wideband power sensors have taken a new approach to the user calibration. The sensor no longer needs to be disconnected from the DUT and connected to a 50 MHz, 0 dBm reference signal — by introducing a known dc reference signal after the RF to dc detection process, all the functions of a user calibration can be achieved, but without the inconvenience. Taking this a step further, transistor switches have been introduced behind the detector to allow a zero to be made while RF is still incident on the RF detector, see Figure 3. This patent-pending approach has been termed as ‘internal zero & cal.’

This capability removes the need for connection and disconnection from the calibration source, thereby reducing test times, measurement uncertainty, and wear and tear on connectors. It is especially useful in manufacturing and automated test environments.

The precision voltage reference device starts with an accuracy of ±3 ppm, and after attenuation and buffering, the stability works out at ±500 ppm (±0.05%). This compares favorably to the 0 dBm, 50 MHz calibrator found on most power meters. For the N1911/2A power meter, the accuracy specification is ±0.4% (25 ± 10 °C) and ±1.2% (0 to 5 °C). Therefore, along with eliminating the associated mismatch uncertainty between the sensor and power reference, there is improved accuracy in the calibration process with this internal calibration.

The N1921/2A power sensors do not rely on the external calibrator on the N1911/2A power meter for calibration (this is included in the power meter to calibrate existing Agilent sensors), so connecting to it offers a means to check the integrity of the detector circuitry and connector. However, this is limited, as it cannot prove or guarantee that the entire operating range of the system is functioning to full accuracy.

What errors can calibration catch? This is best covered with a few examples:

A power sensor may have been exposed to input power that is beyond its absolute maximum rating (damage level), and this could have several effects:

  • The attenuator that is located within the sensor in front of the detector may have failed such that no RF power propagates to the detector. A user calibration or, for the P-series power sensor, a measurement on a known good source (possibly a calibrator) will detect this failure.

  • The actual detector may have been stressed, either to the point of a catastrophic failure, or experiencing accelerated aging. The former effect can be detected by connection to a known source, but the aging effects may be more subtle and are difficult to detect immediately — either from a user calibration or using a known good source.

  • For a thermocouple, if the maximum power is exceeded, the ‘scaling factor’ for the sensor may be modified, and although this can be corrected for by repeating the user calibration, it is likely to reduce the life of the sensor.

The power sensor and meter are used in a variety of situations and environments, some, like system calibrations and installation testing, require several connections and disconnections. The effect of multiple connections can be connector wear and tear. The accuracy impact of this wear and tear manifests itself in two forms:

  1. The VSWR of the sensor will be degraded, causing greater mismatch uncertainty.

  2. The coupling of RF into the detector circuit will be modified, potentially invalidating the applied correction factors. So although the measurement may appear accurate at 50 MHz, with a calibrator, the effect at another frequency may be quite significant.

A calibration kit for a network analyzer is used to transfer connector standards to a reference plane, and care must be taken, not only to not overtighten a connection, but also to ensure that the gauge of the DUT connector is in good shape and will not damage the instrument's connector. The same care needs to be applied with a power sensor in its use to transfer power accuracy to a DUT.

Conclusion

Agilent's internal zero and calibration capability for the power sensor and meter eliminates multiple connections with an external calibration source, with the benefits of reduced measurement uncertainty, reduced connect wear and faster test times. All RF power-sensing technologies rely on some form of correction to achieve optimal accuracy for a power measurement. The user calibration of a sensor, whether using an internal or external source, neither guarantees the validity of the applied corrections, nor the accuracy of the overall measurement. This is achieved when the calibration is used in conjunction with an accurate characterization (when the correction factors are generated) and careful use and handling of the sensor and its RF interface.

References:

  1. The terms in italics are common power measurement specific terms, which are given a fuller definition in App. Note 1449-3, Fundamentals of RF and Microwave Power Measurements (Part 3), Power Measurement Uncertainty per International Guides, Agilent Technologies.

  2. Modern RF sensors are considered as sensors with internal memory (EEPROM) in order to retain Calibration and Correction Factor data, with an extended operating range.

  3. Refer to Measuring Power Levels in Modern Communication Systems at www.home.agilent.com/upload/cmc_upload/All/EPSG090442.pdf also refer to Microwaves & RF, October 2000.

ABOUT THE AUTHOR

Alan B. Anderson is a senior development engineer with Agilent Technologies, whom he joined in 1997 when it was previously Hewlett Packard. He graduated from the University of Glasgow with a European M.Eng. in 1997 and holds an MBA from Edinburgh Business School (2003). At Agilent he has developed its range of peak and wideband RF power sensors. He can be reached at alan_anderson3@agilent.com.



February/March 2012
Part Finder
Search our directory of over 10 million parts.



Popular Searches:
AMP/Tyco Electronics
Maxim Integrated Products
Analog Devices
Molex
Freescale Semiconductor
Advanced Micro Devices
Texas Instruments

 
Back to Top